環境污染/水質分析測量儀器
Environmental Pollution/Water Analysis Instruments
|
|
|
|
NP-500T型激光液體微粒計數儀
|
|
最適宜用于水質純化的激光液體粒度計數和高靈敏度的濁度計,其適合于衛生和勞動福利部臨時政策。非常有用的測量微粒的工具,如在水系統中測量霉菌(采用便攜式型號)。
|
|
This is a highly sensitive
laser particle counter (turbidimeter) for clean
water, most suitable as the counter measurement
against the provisional guide line for clean water
by Ministry of Health, Labor and Welfare. It has a
strong effect on particle counting of
cryptosporidium, etc., jeopardizing safety of tap
water.
Turbidimeter used exclusively for low
turbidity measurement based on the particle counting
method. |
應用領域:
|
|
產品特點:
|
|
同時進行微粒計數和濁度測量,其適合于衛生和勞動福利部臨時政策 |
|
照亮測量區域,無微粒漏檢 |
|
在相同直徑范圍內,能區分霉菌(Chyptosporidium)和鞭毛蟲卵(giardia),數據可以快速得到評估 |
|
可直接聯系到生產線 |
|
可自由設置微粒計數和濁度水平報警限 |
|
光源采用長壽命的半導體激光器 |
|
具有RS-232C和4-20mA(4道)的標準信號輸出 |
|
Both particle counts and turbidity can be measured simultaneously. |
|
Cryptsporidium and giardia are sorted to be the same particle size, so variation in particle size can be judged easily at a glance. |
|
This model can be connected directly to your line of measurement. |
|
You can set an alarm for arbitrary particle counts of turbidity value. |
|
A long life semi-conductor laser is used as a light source. |
|
Contamination detector is equipped with this model. |
|
Data analysis and filing software is available as a standard configuration. |
|
A pump is built in for sample suction and measurement. |
|
|
NP-500T型激光液體微粒計數儀技術參數
|
|
測量范圍 |
微粒計數:7位數字
濁度:0.0001 ~ 2.0mg/I (可切換到NTU) |
測量流量 |
50ml/min |
測量時間 |
微粒數據計算時間可以自由設定,從6秒~99分59秒(也可以無步進設定) |
測量粒徑 |
6通道:0.5µm,1µm,3µm,7µm,12µm,15µm和更大 |
外形尺寸 |
寬度175mm x 深度260mm x 高度240mm |
重量 |
約6.5公斤 |
電源 |
AC電源 |
功率 |
100VA,50/60Hz |
光源 |
半導體激光 |
輸出 |
RS-232C,4-20mAx4道,報警聯系輸出 |
注:此型號的準確度來自于日本顏色研究所(JCRI)的國家標準
|
Dimensions
|
W
175mm
x D
260mm
x H
240mm
|
Weight
|
Approx.
6.5kg
|
Power
supply
|
100V
AC,
50 /
60Hz
|
Power
consumption
|
100VA
|
Light
source
|
Semi-conductor
laser
|
Measuring
items
|
Particle
counts
,
turbidity
|
Measuring
range
|
•
Particle
counts
: 0
to
106
/
50ml
•
Turbidity
:
0.0000º
to
2.0º
(NTU
unit)
|
Measuring
method
|
Lateral
scattering
method
|
Flow
volume
for
measurement
|
50
ml /
min.
|
Measuring
time
|
Elapsed
time
for
particle
data
integration
6
sec.
to 9
min.
59sec.
to
be
arbitrarily
set
(possible
without
interval)
|
Measuring
particle
size
|
6
channels
for
0.5µm
to
1µm,
1µm
to
3µm,
3µm
to
7µm,
7µm
to
12µm,
12µm
to
15µm,
15µm
or
more
|
|
|
|
|